Pulse oximeter bench tests under different simulated skin tones
Swamy, Suvvi K. Narayana; He, Chenyang; Hayes‑Gill, Barrie R.; Hayes-Gill, Barrie R.; Clark, Daniel J.; Green, Sarah; Morgan, Stephen P.
Authors
CHENYANG HE CHENYANG.HE2@NOTTINGHAM.AC.UK
Assistant Professor
BARRIE HAYES-GILL BARRIE.HAYES-GILL@NOTTINGHAM.AC.UK
Professor of Electronic Systems and Medical Devices
Barrie R. Hayes-Gill
Daniel J. Clark
Sarah Green
Prof STEVE MORGAN STEVE.MORGAN@NOTTINGHAM.AC.UK
Professor of Biomedical Engineering
Abstract
Pulse oximeters’ (POs) varying performance based on skin tones has been highly publicised. Compared to arterial blood gas analysis, POs tend to overestimate oxygen saturation (SpO2) values for people with darker skin (occult hypoxemia). The objective is to develop a test bench for assessing commercial home and hospital-based POs in controlled laboratory conditions. A laboratory simulator was used to mimic different SpO2 values (~ 70 to 100%).Different neutral density and synthetic melanin filters were used to reproduce low signal and varying melanin attenuation levels. Six devices consisting of commercial home (Biolight, N = 13; ChoiceMMed, N = 18; MedLinket, N = 9) and hospital-based (Masimo Radical 7 with Neo L, N = 1; GE B450 Masimo SET with LNCS Neo L, N = 1; Nonin 9550 Onyx II™, N = 1) POs were reviewed and their response documented. Significant variations were observed in the recorded SpO2 values among different POs when exposed to identical simulated signals. Differences were greatest for lower SpO2 (< 80%) where empirical data is limited. All PO responses under low signal and melanin attenuation did not change across various simulated SpO2 values. The bench tests do not provide conclusive evidence that melanin does not affect in vivo SpO2 measurements. Research in the areas of instrument calibration, theory and design needs to be further developed. Graphical Abstract: (Figure presented.)
Citation
Swamy, S. K. N., He, C., Hayes‑Gill, B. R., Hayes-Gill, B. R., Clark, D. J., Green, S., & Morgan, S. P. (2024). Pulse oximeter bench tests under different simulated skin tones. Medical and Biological Engineering and Computing, https://doi.org/10.1007/s11517-024-03091-2
Journal Article Type | Article |
---|---|
Acceptance Date | Apr 6, 2024 |
Online Publication Date | Apr 24, 2024 |
Publication Date | Apr 24, 2024 |
Deposit Date | Jun 24, 2024 |
Publicly Available Date | Jun 25, 2024 |
Journal | Medical and Biological Engineering and Computing |
Print ISSN | 0140-0118 |
Electronic ISSN | 1741-0444 |
Publisher | Springer Verlag |
Peer Reviewed | Peer Reviewed |
DOI | https://doi.org/10.1007/s11517-024-03091-2 |
Public URL | https://nottingham-repository.worktribe.com/output/34103750 |
Publisher URL | https://link.springer.com/article/10.1007/s11517-024-03091-2 |
Additional Information | Received: 18 October 2023; Accepted: 6 April 2024; First Online: 24 April 2024 |
Files
Pulse oximeter bench tests
(2 Mb)
PDF
Publisher Licence URL
https://creativecommons.org/licenses/by/4.0/
You might also like
An automated quasi-continuous capillary refill timing device
(2015)
Journal Article
Downloadable Citations
About Repository@Nottingham
Administrator e-mail: discovery-access-systems@nottingham.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search