Skip to main content

Research Repository

Advanced Search

Contrast‐enhanced ultrasound repeatability for the measurement of skeletal muscle microvascular blood flow

Jones, Eleanor J.; Atherton, Philip J.; Piasecki, Mathew; Phillips, Bethan E.

Contrast‐enhanced ultrasound repeatability for the measurement of skeletal muscle microvascular blood flow Thumbnail


Authors

PHILIP ATHERTON philip.atherton@nottingham.ac.uk
Professor of Clinical, metabolic & Molecular Physiology

BETH PHILLIPS beth.phillips@nottingham.ac.uk
Professor of Translational Physiology



Abstract

Contrast-enhanced ultrasound (CEUS) can be used to directly assess skeletal muscle perfusion. However, its repeatability over time has not yet been validated and therefore its use in longitudinal measures (i.e., exploring the impact of a chronic intervention or disease progression) is limited. This study aimed to determine the repeatability of CEUS for the measurement of skeletal muscle microvascular blood flow (MBF) at baseline and in response to exercise, across independent assessment sessions. Ten healthy volunteers (five female; 30 ± 6 years) had CEUS of the right vastus lateralis recorded in two separate sessions, 14 days apart. Measurements were taken at baseline, during an isometric leg extension and during recovery. Acoustic intensity data from a region of interest were plotted as a replenishment curve to obtain blood volume (A) and flow velocity (β) values from a one-phase association non-linear regression of mean tissue echogenicity. Linear regression and Bland–Altman analyses of A and β values were performed, with significance assumed as P < 0.05. Strong positive correlations were observed across sessions for all A and β values (both P < 0.0001). Bland–Altman analysis showed a bias (SD) of −0.013 ± 1.24 for A and −0.014 ± 0.31 for β. A bias of 0.201 ± 0.770 at baseline, 0.527 ± 1.29 during contraction and −0.203 ± 1.29 at recovery was observed for A, and −0.0328 ± 0.0853 (baseline), −0.0446 ± 0.206 (contraction) and 0.0382 ± 0.233 (recovery) for β. A strong agreement between CEUS MBF measures across independent sessions suggests it to be a repeatable method for assessing skeletal muscle perfusion over time, and therefore facilitates wider use in longitudinal studies.

Citation

Jones, E. J., Atherton, P. J., Piasecki, M., & Phillips, B. E. (2023). Contrast‐enhanced ultrasound repeatability for the measurement of skeletal muscle microvascular blood flow. Experimental Physiology, 108(4), 549-553. https://doi.org/10.1113/ep091034

Journal Article Type Article
Acceptance Date Jan 24, 2023
Online Publication Date Feb 4, 2023
Publication Date Apr 1, 2023
Deposit Date Mar 15, 2023
Publicly Available Date Mar 21, 2023
Journal Experimental Physiology
Print ISSN 0958-0670
Electronic ISSN 1469-445X
Publisher Wiley
Peer Reviewed Peer Reviewed
Volume 108
Issue 4
Pages 549-553
DOI https://doi.org/10.1113/ep091034
Keywords blood flow, contrast enhanced ultrasound, repeatability, skeletal muscle
Public URL https://nottingham-repository.worktribe.com/output/17071978
Publisher URL https://physoc.onlinelibrary.wiley.com/doi/10.1113/EP091034

Files





You might also like



Downloadable Citations