Simultaneous in situ temperature and relative humidity monitoring in mechanical ventilators using an array of functionalised optical fibre long period grating sensors

Hromadka, Jiri; Hazlan, Nurul N. Mohd; Hernandez, Francisco U.; Correia, Ricardo; Norris, A.; Morgan, Stephen P; Korposh, Sergiy

Simultaneous in situ temperature and relative humidity monitoring in mechanical ventilators using an array of functionalised optical fibre long period grating sensors Thumbnail


Authors

Jiri Hromadka

Nurul N. Mohd Hazlan

Francisco U. Hernandez

A. Norris

Profile Image

SERHIY KORPOSH S.Korposh@nottingham.ac.uk
Professor of Photonics Instrumentation



Abstract

An array of optical fibre long period gratings (LPGs) has been demonstrated for biomedical application to monitor temperature and relative humidity (RH) changes in the air delivered by a mechanical ventilator operating in different modes. The LPG array consists of two gratings, where one was kept bare to monitor the temperature change and the second was modified with 10 layers of silica nanoparticles to measure relative humidity. A mesoporous film was deposited on the surface of an optical fibre LPG using the layer-by-layer method. The sensor was calibrated in a bench model against a commercially available temperature and relative humidity sensor and sensitivity of the sensor was 0.46 ± 0.01 nm/ oC and 0.53 nm/RH%, respectively. The tip of the sensor array was modified with reflection mirror and placed inside an endotracheal tube (ETT) and tested in typical clinical equipment to enable the in-situ real-time monitoring of humidity and temperature. Temperature and RH changes associated with the breathing frequency of 8 and 15 breaths per minute were successfully monitored using developed sensor array.

Citation

Hromadka, J., Hazlan, N. N. M., Hernandez, F. U., Correia, R., Norris, A., Morgan, S. P., & Korposh, S. (2019). Simultaneous in situ temperature and relative humidity monitoring in mechanical ventilators using an array of functionalised optical fibre long period grating sensors. Sensors and Actuators B: Chemical, 286, 306 - 314. https://doi.org/10.1016/j.snb.2019.01.124

Journal Article Type Article
Acceptance Date Jan 25, 2019
Online Publication Date Jan 28, 2019
Publication Date May 1, 2019
Deposit Date Apr 27, 2020
Publicly Available Date Apr 27, 2020
Journal Sensors and Actuators B: Chemical
Print ISSN 0925-4005
Publisher Elsevier
Peer Reviewed Peer Reviewed
Volume 286
Pages 306 - 314
DOI https://doi.org/10.1016/j.snb.2019.01.124
Keywords Optical fibre sensor, Long period grating (LPG), Healthcare monitoring, Endotracheal tube, Multiplexing, Temperature, Relative humidity (RH), Mechanical ventilation
Public URL https://nottingham-repository.worktribe.com/output/1571168
Publisher URL http://www.sciencedirect.com/science/article/pii/S0925400519301492
Additional Information This article is maintained by: Elsevier; Article Title: Simultaneous in situ temperature and relative humidity monitoring in mechanical ventilators using an array of functionalised optical fibre long period grating sensors; Journal Title: Sensors and Actuators B: Chemical; CrossRef DOI link to publisher maintained version: https://doi.org/10.1016/j.snb.2019.01.124; Content Type: article; Copyright: © 2019 The Authors. Published by Elsevier B.V.