Abstract
Objective: To develop a compact probe that can be used to monitor humidity in ventilator care equipment. A mesoporous film of alternate layers of Poly(allylamine hydrochloride) (PAH) and silica (SiO2) nanoparticles (bilayers), deposited onto an optical fibre was used. The sensing film behaves as a Fabry-Perot cavity of low-finesse where the absorption of water vapour changes the optical thickness and produces a change in reflection proportional to humidity. Methods: The mesoporous film was deposited upon the cleaved tip of an optical fibre using the layer-by-layer method. The sensor was calibrated in a bench model against a commercially available capacitive sensor. The sensitivity and response time were assessed in the range from 5 % relative humidity (RH) to 95 %RH for different numbers of bilayers up to a maximum of nine. Results: The sensitivity increases with the number of bilayers deposited; sensitivity of 2.28 mV/%RH was obtained for nine bilayers. The time constant of the response was 1.13 s ± 0.30 s which is faster than the commercial device (measured as 158 s). After calibration, the optical fibre humidity sensor was utilised in a bench top study employing a mechanical ventilator. The fast response time enabled changes in humidity in individual breaths to be resolved. Conclusion: Optical fibre sensors have the potential to be used to monitor breath to breath humidity during ventilator care. Significance: Control of humidity is an essential part of critical respiratory care and the developed sensor provides a sensitive, compact and fast method of humidity monitoring.
Citation
Hernandez, F. U., Morgan, S. P., Hayes-Gill, B. R., Harvey, D., Kinnear, W., Norris, A., …Korposh, S. (2016). Characterization and use of a fiber optic sensor based on PAH/SiO2 film for humidity sensing in ventilator care equipment. IEEE Transactions on Biomedical Engineering, 63(9), 1985-1992. https://doi.org/10.1109/TBME.2016.2521662
Journal Article Type | Article |
---|---|
Acceptance Date | Oct 1, 2015 |
Online Publication Date | Feb 16, 2016 |
Publication Date | Sep 1, 2016 |
Deposit Date | Sep 12, 2017 |
Publicly Available Date | Sep 12, 2017 |
Journal | IEEE Transactions on Biomedical Engineering |
Print ISSN | 0018-9294 |
Electronic ISSN | 1558-2531 |
Publisher | Institute of Electrical and Electronics Engineers |
Peer Reviewed | Peer Reviewed |
Volume | 63 |
Issue | 9 |
Pages | 1985-1992 |
DOI | https://doi.org/10.1109/TBME.2016.2521662 |
Keywords | Humidity, Optical fibers, Films, Humidity measurement, Biomedical measurement, Probes, Time factors |
Public URL | https://nottingham-repository.worktribe.com/output/975095 |
Publisher URL | http://ieeexplore.ieee.org/document/7393540/ |
Additional Information | (c) 2016 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other users, including reprinting/ republishing this material for advertising or promotional purposes, creating new collective works for resale or redistribution to servers or lists, or reuse of any copyrighted components of this work in other works. |
Contract Date | Sep 12, 2017 |
Files
TBME-01015-2015_revised_markedFUH-final.pdf
(1 Mb)
PDF
You might also like
An automated quasi-continuous capillary refill timing device
(2015)
Journal Article
A hybrid neural network/rule-based technique for on-line gesture and hand-written character recognition
(-0001)
Presentation / Conference Contribution
Downloadable Citations
About Repository@Nottingham
Administrator e-mail: discovery-access-systems@nottingham.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search