Abstract

C-reactive protein (CRP) is one of the most expressed proteins in blood during acute phase inflammation, and its minute level increase has also been recognized for the clinical diagnosis of cardio vascular diseases. Unfortunately, the available commercial immunoassays are labour intensive, require large sample volumes, and have practical limitations, such as low stability and high production costs. Hence, we have developed a simple, cost effective, and label-free electrochemical immunoassay for the measurement of CRP in a drop of serum sample using an immunosensor strip made up of a screen printed carbon electrode (SPE) modified with anti-CRP functionalized gold nanoparticles (AuNPs). The measurement relies on the decrease of the oxidation current of the redox indicator Fe3+/Fe2+, resulting from the immunoreaction between CRP and anti-CRP. Under optimal conditions, the present immunoassay measures CRP in a linear range from 0.4–200 nM (0.047–23.6 µg mL−1), with a detection limit of 0.15 nM (17 ng mL−1, S/N = 3) and sensitivity of 90.7 nA nM−1, in addition to a good reproducibility and storage stability. The analytical applicability of the presented immunoassay is verified by CRP measurements in human blood serum samples. This work provides the basis for a low-priced, safe, and easy-to-use point-of-care immunosensor assay to measure CRP at clinically relevant concentrations.

Citation

Thangamuthu, M., Santschi, C., & J. F. Martin, O. (2018). Label-Free Electrochemical Immunoassay for C-Reactive Protein. Biosensors, 8(2), Article 34. https://doi.org/10.3390/bios8020034

Journal Article Type Article
Acceptance Date Mar 27, 2018
Online Publication Date Mar 30, 2018
Publication Date 2018-06
Deposit Date Aug 1, 2023
Publicly Available Date Aug 4, 2023
Journal Biosensors
Publisher MDPI
Peer Reviewed Peer Reviewed
Volume 8
Issue 2
Article Number 34
DOI https://doi.org/10.3390/bios8020034
Keywords screen printed electrode (SPE); C-reactive protein; electrochemical assay; label-free Immunosensor; gold nanoparticles