Abstract

Capillary refill time (CRT) refers to the time taken for body tissue to regain its colour after an applied blanching pressure is released. Usually, pressure is manually applied and not measured. Upon release of pressure, simple mental counting is typically used to estimate how long it takes for the skin to regain its colour. However, this method is subjective and can provide inaccurate readings due to human error. CRT is often used to assess shock and hydration but also has the potential to assess peripheral arterial disease which can result in tissue breakdown, foot ulcers and ultimately amputation, especially in people with diabetes. The aim of this study was to design an optical fibre sensor to simultaneously detect blood volume changes and the contact pressure applied to the foot. The CRT probe combines two sensors: a plastic optical fibre (POF) based on photoplethysmography (PPG) to measure blood volume changes and a fibre Bragg grating to measure skin contact pressure. The results from 10 healthy volunteers demonstrate that the blanching pressure on the subject’s first metatarsal head of the foot was 100.8 ± 4.8 kPa (mean and standard deviation), the average CRT was 1.37 ± 0.46 s and the time to achieve a stable blood volume was 4.77 ± 1.57 s. For individual volunteers, the fastest CRT measured was 0.82 ± 0.11 and the slowest 1.94 ± 0.49 s. The combined sensor and curve fitting process has the potential to provide increased reliability and accuracy for CRT measurement of the foot in diabetic foot ulcer clinics and in the community.

Citation

Ballaji, H. K., Correia, R., Liu, C., Korposh, S., Hayes-Gill, B. R., Musgrove, A., & Morgan, S. P. (2021). Optical Fibre Sensor for Capillary Refill Time and Contact Pressure Measurements under the Foot. Sensors, 21(18), Article 6072. https://doi.org/10.3390/s21186072

Journal Article Type Article
Acceptance Date Aug 20, 2021
Online Publication Date Sep 10, 2021
Publication Date Sep 1, 2021
Deposit Date Sep 6, 2021
Publicly Available Date Sep 6, 2021
Journal Sensors
Electronic ISSN 1424-8220
Publisher MDPI
Peer Reviewed Peer Reviewed
Volume 21
Issue 18
Article Number 6072
DOI https://doi.org/10.3390/s21186072
Keywords Electrical and Electronic Engineering; Biochemistry; Instrumentation; Atomic and Molecular Physics, and Optics; Analytical Chemistry
Public URL https://nottingham-repository.worktribe.com/output/6184858
Publisher URL https://www.mdpi.com/1424-8220/21/18/6072/htm

Files





You might also like



Downloadable Citations