Y. Msayib
Robust estimation of quantitative perfusion from multi‐phase pseudo‐continuous arterial spin labeling
Msayib, Y.; Craig, M.; Simard, M. A.; Larkin, J. R.; Shin, D. D.; Liu, T. T.; Sibson, N. R.; Okell, T. W.; Chappell, M. A.
Authors
M. Craig
M. A. Simard
J. R. Larkin
D. D. Shin
T. T. Liu
N. R. Sibson
T. W. Okell
Prof MICHAEL CHAPPELL MICHAEL.CHAPPELL@NOTTINGHAM.AC.UK
Professor of Biomedical Imaging
Abstract
Purpose
Multi‐phase PCASL has been proposed as a means to achieve accurate perfusion quantification that is robust to imperfect shim in the labeling plane. However, there exists a bias in the estimation process that is a function of noise in the data. In this work, this bias is characterized and then addressed in animal and human data.
Methods
The proposed algorithm to overcome bias uses the initial biased voxel‐wise estimate of phase tracking error to cluster regions with different off‐resonance phase shifts, from which a high‐SNR estimate of regional phase offset is derived. Simulations were used to predict the bias expected at typical SNR. Multi‐phase PCASL in 3 rat strains (n = 21) at 9.4 T was considered, along with 20 human subjects previously imaged using ASL at 3 T. The algorithm was extended to include estimation of arterial blood flow velocity.
Results
Based on simulations, a perfusion estimation bias of 6‐8% was expected using 8‐phase data at typical SNR. This bias was eliminated when a high‐precision estimate of phase error was available. In the preclinical data, the bias‐corrected measure of perfusion (107 ± 14 mL/100g/min) was lower than the standard analysis (116 ± 14 mL/100g/min), corresponding to a mean observed bias across strains of 8.0%. In the human data, bias correction resulted in a 15% decrease in the estimate of perfusion.
Conclusions
Using a retrospective algorithmic approach, it was possible to exploit common information found in multiple voxels within a whole region of the brain, offering superior SNR and thus overcoming the bias in perfusion quantification from multi‐phase PCASL.
Citation
Msayib, Y., Craig, M., Simard, M. A., Larkin, J. R., Shin, D. D., Liu, T. T., …Chappell, M. A. (2020). Robust estimation of quantitative perfusion from multi‐phase pseudo‐continuous arterial spin labeling. Magnetic Resonance in Medicine, 83(3), 815-829. https://doi.org/10.1002/mrm.27965
Journal Article Type | Article |
---|---|
Acceptance Date | Aug 2, 2019 |
Online Publication Date | Aug 20, 2019 |
Publication Date | 2020-03 |
Deposit Date | Sep 28, 2020 |
Publicly Available Date | Oct 15, 2020 |
Journal | Magnetic Resonance in Medicine |
Print ISSN | 0740-3194 |
Electronic ISSN | 1522-2594 |
Publisher | Wiley |
Peer Reviewed | Peer Reviewed |
Volume | 83 |
Issue | 3 |
Pages | 815-829 |
DOI | https://doi.org/10.1002/mrm.27965 |
Keywords | Radiology Nuclear Medicine and imaging |
Public URL | https://nottingham-repository.worktribe.com/output/4930898 |
Publisher URL | https://onlinelibrary.wiley.com/doi/full/10.1002/mrm.27965 |
Files
Robust estimation of quantitative perfusion
(1.7 Mb)
PDF
Publisher Licence URL
https://creativecommons.org/licenses/by/4.0/
You might also like
BASIL: A Toolbox for Perfusion Quantification using Arterial Spin Labelling
(2023)
Journal Article
Unified Surface and Volumetric Inference on Functional Imaging Data
(2023)
Presentation / Conference Contribution
Downloadable Citations
About Repository@Nottingham
Administrator e-mail: discovery-access-systems@nottingham.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search