Across‐vendor standardization of semi‐LASER for single‐voxel MRS at 3T
Kantarci, Kejal; Deelchand, Dinesh K.; Berrington, Adam; Noeske, Ralph; Joers, James M.; Arani, Arvin; Gillen, Joseph; Sch�r, Michael; Nielsen, Jon-Fredrik; Peltier, Scott; Seraji-Bozorgzad, Navid; Landheer, Karl; Juchem, Christoph; Soher, Brian J.; Noll, Douglas C.; Ratai, Eva M.; Mareci, Thomas H.; Barker, Peter B.; �z, G�lin
Authors
Dinesh K. Deelchand
Dr ADAM BERRINGTON ADAM.BERRINGTON@NOTTINGHAM.AC.UK
RESEARCH FELLOW
Ralph Noeske
James M. Joers
Arvin Arani
Joseph Gillen
Michael Sch�r
Jon-Fredrik Nielsen
Scott Peltier
Navid Seraji-Bozorgzad
Karl Landheer
Christoph Juchem
Brian J. Soher
Douglas C. Noll
Eva M. Ratai
Thomas H. Mareci
Peter B. Barker
G�lin �z
Abstract
The semi‐adiabatic localization by adiabatic selective refocusing (sLASER) sequence provides single‐shot full intensity signal with clean localization and minimal chemical shift displacement error and was recommended by the international MRS Consensus Group as the preferred localization sequence at high‐ and ultra‐high fields. Across‐vendor standardization of the sLASER sequence at 3 tesla has been challenging due to the B1 requirements of the adiabatic inversion pulses and maximum B1 limitations on some platforms. The aims of this study were to design a short‐echo sLASER sequence that can be executed within a B1 limit of 15 μT by taking advantage of gradient‐modulated RF pulses, to implement it on three major platforms and to evaluate the between‐vendor reproducibility of its perfomance with phantoms and in vivo. In addition, voxel‐based first and second order B0 shimming and voxel‐based B1 adjustments of RF pulses were implemented on all platforms. Amongst the gradient‐modulated pulses considered (GOIA, FOCI and BASSI), GOIA‐WURST was identified as the optimal refocusing pulse that provides good voxel selection within a maximum B1 of 15 μT based on localization efficiency, contamination error and ripple artifacts of the inversion profile. An sLASER sequence (30 ms echo time) that incorporates VAPOR water suppression and 3D outer volume suppression was implemented with identical parameters (RF pulse type and duration, spoiler gradients and inter‐pulse delays) on GE, Philips and Siemens and generated identical spectra on the GE ‘Braino’ phantom between vendors. High‐quality spectra were consistently obtained in multiple regions (cerebellar white matter, hippocampus, pons, posterior cingulate cortex and putamen) in the human brain across vendors (5 subjects scanned per vendor per region; mean signal‐to‐noise ratio [less than] 33; mean water linewidth between 6.5 Hz to 11.4 Hz). The harmonized sLASER protocol is expected to produce high reproducibility of MRS across sites thereby allowing large multi‐site studies with clinical cohorts.
Citation
Kantarci, K., Deelchand, D. K., Berrington, A., Noeske, R., Joers, J. M., Arani, A., Gillen, J., Schär, M., Nielsen, J.-F., Peltier, S., Seraji-Bozorgzad, N., Landheer, K., Juchem, C., Soher, B. J., Noll, D. C., Ratai, E. M., Mareci, T. H., Barker, P. B., & Öz, G. (2021). Across‐vendor standardization of semi‐LASER for single‐voxel MRS at 3T. NMR in Biomedicine, 34(5), Article e4218. https://doi.org/10.1002/nbm.4218
Journal Article Type | Article |
---|---|
Acceptance Date | Oct 21, 2019 |
Online Publication Date | Dec 18, 2019 |
Publication Date | 2021-05 |
Deposit Date | Jan 3, 2020 |
Publicly Available Date | Dec 19, 2020 |
Journal | NMR in Biomedicine |
Print ISSN | 0952-3480 |
Electronic ISSN | 1099-1492 |
Publisher | Wiley |
Peer Reviewed | Peer Reviewed |
Volume | 34 |
Issue | 5 |
Article Number | e4218 |
DOI | https://doi.org/10.1002/nbm.4218 |
Keywords | Brain, Clinical, GOIA-WURST, Gradient-modulated, Human, Harmonization, MR spectroscopy |
Public URL | https://nottingham-repository.worktribe.com/output/3659046 |
Publisher URL | https://onlinelibrary.wiley.com/doi/full/10.1002/nbm.4218 |
Additional Information | This is the peer reviewed version of the following article: Deelchand, DK, Berrington, A, Noeske, R, et al. Across‐vendor standardization of semi‐LASER for single‐voxel MRS at 3T. NMR in Biomedicine. 2019;e4218, which has been published in final form at https://doi.org/10.1002/nbm.4218. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Use of Self-Archived Versions. |
Files
SLASER Standardization R2 Accepted
(967 Kb)
PDF
You might also like
Calibration-free regional RF shims for MRS
(2021)
Journal Article
Downloadable Citations
About Repository@Nottingham
Administrator e-mail: discovery-access-systems@nottingham.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2025
Advanced Search