Glyn S. Spencer
Exploring the origins of EEG motion artefacts during simultaneous fMRI acquisition: implications for motion artefact correction
Spencer, Glyn S.; Smith, James A.; Chowdhury, Muhammad E.H.; Bowtell, Richard W.; Mullinger, Karen J.
Authors
James A. Smith
Muhammad E.H. Chowdhury
Professor RICHARD BOWTELL RICHARD.BOWTELL@NOTTINGHAM.AC.UK
Professor of Physics
Dr KAREN MULLINGER KAREN.MULLINGER@NOTTINGHAM.AC.UK
Associate Professor
Abstract
Motion artefacts (MAs) are induced within EEG data collected simultaneously with fMRI when the subject’s head rotates relative to the magnetic field. The effects of these artefacts have generally been ameliorated by removing periods of data during which large artefact voltages appear in the EEG traces. However, even when combined with other standard post-processing methods, this strategy does not remove smaller MAs which can dominate the neuronal signals of interest. A number of methods are therefore being developed to characterise the MA by measuring reference signals and then using these in artefact correction. These methods generally assume that the head and EEG cap, plus any attached sensors, form a rigid body which can be characterised by a standard set of six motion parameters. Here we investigate the motion of the head/EEG cap system to provide a better understanding of MAs. We focus on the reference layer artefact subtraction (RLAS) approach, as this allows measurement of a separate reference signal for each electrode that is being used to measure brain activity.
Through a series of experiments on phantoms and subjects, we find that movement of the EEG cap relative to the phantom and skin on the forehead is relatively small and that this non-rigid body movement does not appear to cause considerable discrepancy in artefacts between the scalp and reference signals. However, differences in the amplitude of these signals is observed which may be due to differences in geometry of the system from which the reference signals are measured compared with the brain signals. In addition, we find that there is non-rigid body movement of the skull and skin which produces an additional MA component for a head shake, which is not present for a head nod. This results in a large discrepancy in the amplitude and temporal profile of the MA measured on the scalp and reference layer, reducing the efficacy of MA correction based on the reference signals.
Together our data suggest that the efficacy of the correction of MA using any reference-based system is likely to differ for different types of head movement with head shake being the hardest to correct. This provides new information to inform the development of hardware and post-processing methods for removing MAs from EEG data acquired simultaneously with fMRI data.
Citation
Spencer, G. S., Smith, J. A., Chowdhury, M. E., Bowtell, R. W., & Mullinger, K. J. (in press). Exploring the origins of EEG motion artefacts during simultaneous fMRI acquisition: implications for motion artefact correction. NeuroImage, 173, https://doi.org/10.1016/j.neuroimage.2018.02.034
Journal Article Type | Article |
---|---|
Acceptance Date | Feb 16, 2018 |
Online Publication Date | Feb 25, 2018 |
Deposit Date | Feb 26, 2018 |
Publicly Available Date | Feb 26, 2018 |
Journal | NeuroImage |
Print ISSN | 1053-8119 |
Electronic ISSN | 1095-9572 |
Publisher | Elsevier |
Peer Reviewed | Peer Reviewed |
Volume | 173 |
DOI | https://doi.org/10.1016/j.neuroimage.2018.02.034 |
Keywords | Simultaneous EEG-fMRI; Artefact correction; Motion artefacts; Motion characterisation |
Public URL | https://nottingham-repository.worktribe.com/output/916815 |
Publisher URL | https://www.sciencedirect.com/science/article/pii/S1053811918301319 |
Files
1-s2.0-S1053811918301319-main.pdf
(1.7 Mb)
PDF
Copyright Statement
Copyright information regarding this work can be found at the following address: http://creativecommons.org/licenses/by/4.0
You might also like
Deuterium brain imaging at 7T during D2O dosing
(2022)
Journal Article
Improved Dynamic Contrast-Enhanced MRI Using Low Rank with Joint Sparsity
(2022)
Journal Article
A lightweight magnetically shielded room with active shielding
(2022)
Journal Article
Downloadable Citations
About Repository@Nottingham
Administrator e-mail: digital-library-support@nottingham.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search