Daly, Susan; Graham, Peter A.; Freeman, Kathleen P.

Quality control validation for a veterinary laboratory network of six Sysmex XT-2000i/V hematology analyzers iV hematology analyzers Thumbnail


Authors

Susan Daly

Dr PETER GRAHAM PETER.GRAHAM@NOTTINGHAM.AC.UK
Professor of Veterinary Clinical Pathology and Endocrinology

Kathleen P. Freeman



Abstract

Background: Quality control (QC) validation is an important step in the laboratory harmonization process. This includes the application of statistical QC requirements, procedures, and control rules to identify and maintain ongoing stable analytical performance. This provides confidence in the production of patient results that are suitable for clinical interpretation across a network of veterinary laboratories. Objectives: To determine that a higher probability of error detection (Ped) and lower probability of false rejection (Pfr) using a simple control rule and one level of quality control material (QCM) could be achieved using observed analytical performance than by using the manufacturer's acceptable ranges for QCM on the Sysmex XT-2000iV hematology analyzers for veterinary use. We also determined whether Westgard Sigma Rules could be sufficient to monitor and maintain a sufficiently high level of analytical performance to support harmonization. Methods: EZRules3 was used to investigate candidate QC rules and determine the Ped and Pfr of manufacturer's acceptable limits and also analyzer-specific observed analytical performance for each of the six Sysmex analyzers within our laboratory system using the American Society of Veterinary Clinical Pathology (ASVCP)-recommended or internal expert opinion quality goals (expressed as total allowable error, TEa) as the quality requirement. The internal expert quality goals were generated by consensus of the Quality, Education, Planning, and Implementation (QEPI) group comprised of five clinical pathologists and seven laboratory technicians and managers. Sigma metrics, which are a useful monitoring tool and can be used in conjunction with Westgard Sigma Rules, were also calculated. Results: The QC validation using the manufacturer's acceptable limits for analyzer 1 showed only 3/10 measurands reached acceptable Ped for veterinary laboratories (>0.85). For QC validation based on observed analyzer performance, the Ped was >0.94 using a 1-2.5s QC rule for the majority of observations (57/60) across the group of analyzers at the recommended TEa. We found little variation in Pfr between manufacturer acceptable limits and individual analyzer observed performance as this is a characteristic of the rule used, not the analyzer performance. Conclusions: An improved probability of error detection and probability of false rejection using a 1-2.5s QC rule for individual analyzer QC was achieved compared with the use of the manufacturers' acceptable limits for hematology in veterinary laboratories. A validated QC rule (1-2.5s) in conjunction with sigma metrics (>5.5), desirable bias, and desirable CV based on biologic variation was successful to evaluate stable analytical performance supporting continued harmonization across the network of analyzers.

Citation

Daly, S., Graham, P. A., & Freeman, K. P. (2022). Quality control validation for a veterinary laboratory network of six Sysmex XT-2000i/V hematology analyzers iV hematology analyzers. Veterinary Clinical Pathology, 51(4), 565-576. https://doi.org/10.1111/vcp.13163

Journal Article Type Article
Acceptance Date May 10, 2022
Online Publication Date Aug 3, 2022
Publication Date 2022-12
Deposit Date May 20, 2022
Publicly Available Date Aug 12, 2022
Journal Veterinary Clinical Pathology
Print ISSN 0275-6382
Electronic ISSN 1939-165X
Publisher Wiley
Peer Reviewed Peer Reviewed
Volume 51
Issue 4
Pages 565-576
DOI https://doi.org/10.1111/vcp.13163
Keywords General Veterinary
Public URL https://nottingham-repository.worktribe.com/output/8136711
Publisher URL https://onlinelibrary.wiley.com/doi/10.1111/vcp.13163

Files