Diwei He
A single-chip CMOS pulse oximeter with on-chip lock-in detection
He, Diwei; Morgan, Stephen P.; Trachanis, Dimitrios; van Hese, Jan; Drogoudis, Dimitris; Fummi, Franco; Stefanni, Francesco; Guarnieri, Valerio; Hayes-Gill, Barrie R.
Authors
Prof STEVE MORGAN STEVE.MORGAN@NOTTINGHAM.AC.UK
Professor of Biomedical Engineering
Dimitrios Trachanis
Jan van Hese
Dimitris Drogoudis
Franco Fummi
Francesco Stefanni
Valerio Guarnieri
BARRIE HAYES-GILL BARRIE.HAYES-GILL@NOTTINGHAM.AC.UK
Professor of Electronic Systems and Medical Devices
Abstract
Pulse oximetry is a noninvasive and continuous method for monitoring the blood oxygen saturation level. This paper presents the design and testing of a single-chip pulse oximeter fabricated in a 0.35 µm CMOS process. The chip includes photodiode, transimpedance amplifier, analogue band-pass filters, analogue-to-digital converters, digital signal processor and LED timing control. The experimentally measured AC and DC characteristics of individual circuits including the DC output voltage of the transimpedance amplifier, transimpedance gain of the transimpedance amplifier, and the central frequency and bandwidth of the analogue band-pass filters, show a good match (within 1%) with the circuit simulations. With modulated light source and integrated lock-in detection the sensor effectively suppresses the interference from ambient light and 1/f noise. In a breath hold and release experiment the single chip sensor demonstrates consistent and comparable performance to commercial pulse oximetry devices with a mean of 1.2% difference. The single-chip sensor enables a compact and robust design solution that offers a route towards wearable devices for health monitoring
Citation
He, D., Morgan, S. P., Trachanis, D., van Hese, J., Drogoudis, D., Fummi, F., Stefanni, F., Guarnieri, V., & Hayes-Gill, B. R. (2015). A single-chip CMOS pulse oximeter with on-chip lock-in detection. Sensors, 15(7), 17076-17088. https://doi.org/10.3390/s150717076
Journal Article Type | Article |
---|---|
Acceptance Date | Jul 6, 2015 |
Online Publication Date | Jul 14, 2015 |
Publication Date | Jul 14, 2015 |
Deposit Date | Sep 15, 2017 |
Publicly Available Date | Feb 6, 2019 |
Electronic ISSN | 1424-8220 |
Publisher | MDPI |
Peer Reviewed | Peer Reviewed |
Volume | 15 |
Issue | 7 |
Pages | 17076-17088 |
DOI | https://doi.org/10.3390/s150717076 |
Public URL | https://nottingham-repository.worktribe.com/output/1105680 |
Publisher URL | https://www.mdpi.com/1424-8220/15/7/17076 |
Related Public URLs | http://www.scopus.com/inward/record.url?eid=2-s2.0-84940200241&partnerID=40&md5=1fefe61c63ae0e5caf13e6ea3f38ec57 |
Contract Date | Feb 6, 2019 |
Files
A Single-Chip CMOS
(1.4 Mb)
PDF
Publisher Licence URL
https://creativecommons.org/licenses/by/4.0/
You might also like
Pulse oximeter bench tests under different simulated skin tones
(2024)
Journal Article
Forehead monitoring of heart rate in neonatal intensive care
(2023)
Journal Article
Downloadable Citations
About Repository@Nottingham
Administrator e-mail: discovery-access-systems@nottingham.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search