Reducing the gradient artefact in simultaneous EEG-fMRI by adjusting the subject’s axial position

Mullinger, Karen J.; Yan, Winston X.; Bowtell, Richard W.

Authors

Winston X. Yan



Abstract

Large artefacts which compromise EEG data quality are generated when electroencephalography (EEG) and functional magnetic resonance imaging (fMRI) are carried out concurrently. The gradient artefact produced by the time-varying magnetic field gradients is the largest of these artefacts. Although average artefact correction (AAS) and related techniques can remove the majority of this artefact, the need to avoid amplifier saturation necessitates the use of a large dynamic range and strong low-pass filtering in EEG recording. Any intrinsic reduction in the gradient artefact amplitude would allow data with a higher bandwidth to be acquired without amplifier saturation, thus increasing the frequency range of neuronal activity that can be investigated using combined EEG-fMRI. Furthermore, gradient artefact correction methods assume a constant artefact morphology over time, so their performance is compromised by subject movement. Since the resulting, residual gradient artefacts can easily swamp signals from brain activity, any reduction in their amplitude would be highly advantageous for simultaneous EEG-fMRI studies. The aim of this work was to investigate whether adjustment of the subject’s axial position in the MRI scanner can reduce the amplitude of the induced gradient artefact, before and after artefact correction using AAS. The variation in gradient artefact amplitude as a function of the subject’s axial position was first investigated in six subjects by applying gradient pulses along the three Cartesian axes. The results of this study showed that a significant reduction in the gradient artefact magnitude can be achieved by shifting the subject axially by 4 cm towards the feet relative to the standard subject position (nasion at iso-centre). In a further study, the 4 cm shift was shown to produce a 40% reduction in the RMS amplitude (and a 31% reduction in the range) of the gradient artefact generated during the execution of a standard multi-slice, EPI sequence. By picking out signals occurring at harmonics of the slice acquisition frequency, it was also shown that the 4 cm shift led to a 36% reduction in the residual gradient artefact after AAS. Functional and anatomical MR data quality is not affected by the 4 cm shift, as the head remains in the homogeneous region of the static magnet field and gradients.

Citation

Mullinger, K. J., Yan, W. X., & Bowtell, R. W. (2011). Reducing the gradient artefact in simultaneous EEG-fMRI by adjusting the subject’s axial position. NeuroImage, 54(3), https://doi.org/10.1016/j.neuroimage.2010.09.079

Journal Article Type Article
Publication Date Jan 1, 2011
Deposit Date Mar 25, 2011
Publicly Available Date Mar 25, 2011
Journal NeuroImage
Print ISSN 1053-8119
Electronic ISSN 1053-8119
Publisher Elsevier
Peer Reviewed Peer Reviewed
Volume 54
Issue 3
DOI https://doi.org/10.1016/j.neuroimage.2010.09.079
Public URL https://nottingham-repository.worktribe.com/output/1011107
Publisher URL http://dx.doi.org/10.1016/j.neuroimage.2010.09.079

Files





You might also like



Downloadable Citations